In short
A food producer's microbiology lab tests sugar solutions for microbes. Each test pulls a measured amount of sample through a thin membrane filter under vacuum. The microbes stay on the membrane, the membrane is laid onto a dish of growth medium (an agar plate), and the plate goes into an incubator so that any microbes grow into countable colonies.
Today all of it is done by hand. Every handling step is a chance to contaminate the sample, and contamination shows up as a false result. The lab wants a system that runs the whole sequence, stays clean between samples, and records what happened to every sample.
| What | How much |
|---|---|
| Samples | up to 100 a day, in one 8-hour shift |
| Filtrations | 7 per sample (2 boiled, 5 unboiled): 700 a day |
| Time per filtration | about 40 seconds on average |
| Environment | a semi-sterile enclosure, not a cleanroom |
What one sample turns into
Every sugar solution is tested twice over: a boiled portion and an unboiled one. The boiled portion goes onto two plates, the unboiled one onto five. V1 to V4 are four different agar media, each with its own incubation.
| Portion | Medium | Plates | Sample per plate | Incubation |
|---|---|---|---|---|
| Boiled | V1 | 2 | 15 ml | 48 h at 55 °C |
| Unboiled | V2 | 2 | 3 ml | 48 h at 30 °C |
| Unboiled | V3 | 2 | 15 ml | 72 h at 30 °C |
| Unboiled | V4 | 1 | 40 ml | 48 h at 36 °C |
Every sample volume goes on top of 100 ml of sterile water and is followed by a 100 ml wash, so up to about 240 ml passes through the funnel per filtration.
How one sample is processed
The lab breaks the work into 19 steps. Steps 5 to 16 repeat for every membrane, seven times per sample.
Before a run
- Disinfect the system. The method is still open: see Keeping it sterile.
- The operator loads sterile frits, funnels, membrane filters, pipette tips and agar plates.
- The operator provides sterile process water, a liquid waste container and waste bags.
- Prepared, sterile samples come in through a defined hand-over point.
Start of the sample
- 1Take the sample, its boiled and its unboiled portion.
- 2Scan its QR code. The code selects the protocol: volumes, number of filtrations, media and incubation.
- 3Open it. The lab's step diagram takes the lid off here and puts it back at the end; a septum or pierceable lid is still an option.
- 4Put a frit on the filter base. It stays for all seven membranes.
Every membrane, seven times
- 5Membrane on, from the dispenser onto the frit.
- 6Funnel on. A sterile funnel locks onto the filter block.
- 7Water in: 100 ml of sterile process water.
- 8Dose and filter. A fresh sterile tip doses the sample volume (3, 15 or 40 ml). The vacuum valve opens, a pressure or liquid sensor detects the end of filtration, and another 100 ml of sterile water washes through.
- 9Tip to waste.
- 10Funnel off.
- 11Scan the agar plate that belongs to this membrane.
- 12Open the plate.
- 13Lift the membrane off the frit.
- 14Lay it on the agar, flat, without folds or trapped air.
- 15Close the plate.
- 16Plate to its magazine, one per incubation condition.
End of the sample
- 17Frit off for cleaning, or a disinfection step instead.
- 18Close the sample.
- 19Sample to its magazine.
Keeping it sterile
Every plate is a test: whatever the system carries from one sample to the next grows on the agar and reads as a false result. The negative controls exist to catch exactly that. So the parts that touch every sample (grippers, frits, funnels, filtration units) have to be clean before the next sample starts.
The lab has already ruled some methods out, and the team behind the TU Dresden demonstrator (next section) suggested others:
| Method | What is known | From |
|---|---|---|
| UV light | Not enough on its own: shadowed surfaces stay untreated | the lab |
| Hydrogen peroxide (H₂O₂) | Expensive | the lab |
| Flammable spray, e.g. alcohol | Expensive: it needs extraction so no explosive atmosphere forms | the lab |
| Hot air | What the TU Dresden demonstrator uses | TU Dresden |
| Steam, sterilising in place | 121 °C plus a holding time; the system has to run under overpressure | TU Dresden |
| Ozone | Possible; afterwards an activated-carbon filter breaks it down to oxygen | TU Dresden |
| Infrared, ultrasound | Not assessed yet | — |
Figure 4Disinfection between samples
Still to settle: which method fits inside the cycle time, whether it runs automatically between samples, and whether a manual ethanol wipe before production starts is still needed.
The TU Dresden demonstrator
At analytica 2026 in Munich, TU Dresden and its partners showed smartAQAlab, a partly automated system for microbiological and chemical water analysis. It automates the two steps at the heart of this use case, membrane filtration and laying the membrane onto agar, but for drinking and process water rather than sugar solutions. The lab has looked at it as a reference.
- Status: still in development in Dresden; the project is not finished.
- Focus: raising throughput.
- Sterilisation: hot air.
- Who: the chair of bioprocess engineering at TU Dresden, with MLE, SmartLab Solutions and INTEGRIS LIMS, co-funded by the EU and the Free State of Saxony (project page).
What it shows: a gripper placing a membrane on a frit, a filtration unit, and tweezers laying the filtered membrane flat on an agar plate. What it does not show yet: published throughput figures, or seven membranes per sample at about 40 seconds each.
Where the system sits
Consumables and storage
Funnels, agar plates, membrane filters, pipette tips and frits are stored on board and presented automatically: a carousel, stack magazines or flexible stacking systems. Finished plates need buffered storage, kept apart by incubation condition, until they go into the incubators. At 100 samples a day:
| Incubation | Medium | Plates a day |
|---|---|---|
| 48 h at 55 °C | V1 | 200 |
| 48 h at 30 °C | V2 | 200 |
| 72 h at 30 °C | V3 | 200 |
| 48 h at 36 °C | V4 | 100 |
Water supply and waste
- Supply: sterile process water in bag systems, with feed pumps and dosing.
- Waste: a liquid waste container, automatic disposal of solid waste, replaceable waste bags.
Interfaces
- Upstream, which may be automated later: dissolving the samples, preparing the media, pouring the agar plates.
- Downstream: moving finished plates into the incubators, recording the data, tracking every sample by QR code.
Traceability
Every step is recorded against the sample: sample number, filtration parameters, the agar plate used, timestamps and every operator intervention.
Requirements
Must have · 13
- ≤ 100 samples/day
throughput
Up to 100 samples in one 8-hour shift
- ≤ 700 filtrations/day
filtrations
7 filtrations per sample (2 boiled, 5 unboiled): up to 700 a day
- 40 s
cycle time
About 40 seconds per filtration on average (480 minutes / 700 filtrations), so about 4.8 minutes per sample
- 10 %
dosing tolerance
Pipetting within the ±10% the lab accepts (the instrument spec is ±2%). Over-dosing is tolerated; under-dosing is not
sterility
No carry-over between samples: everything a sample touches is cleaned, swapped or disinfected before the next one
processing order
Within a sample, the 2 boiled filtrations run before the 5 unboiled ones, all on the same frit
traceability
Every step recorded against the sample: sample number, filtration parameters, the plate used (by its QR code), timestamps, and every operator intervention
protocol
The filtration protocol is chosen per sample from its QR code: volume, number of filtrations, medium, incubation
- 3–40 ml
dosing range
Doses sample volumes of 3, 15 and 40 ml on top of 100 ml sterile water, followed by a 100 ml sterile-water wash
environment
Runs semi-sterile inside an enclosure, not in a cleanroom. The process is not GMP
consumables
Sterile funnels, membranes, frits, pipette tips and agar plates are stored on board and presented automatically
plate handling
Finished plates are buffered and kept apart by incubation condition (48 h at 55 °C, 48 h at 30 °C, 72 h at 30 °C, 48 h at 36 °C) until they go to the incubators
supply and waste
Sterile process water from bag systems with pumps and dosing; liquid waste container; automatic solid-waste disposal into replaceable bags
Should have · 3
- ≥ 120 min
walk-away
At least 120 minutes running unattended
- ≤ 10 min
fault handling
A filtration still running after 10 minutes is detected, alarmed and recovered, and the run continues
controls
Negative controls with sterile water at the start, middle and end of each run
What is still open
The biggest open choice is the robot concept: an arm with a tool changer (gripper, suction, pipetting module, membrane handling), or a gantry with fixed tools that moves in height and width, needs no tool changes and can run more work in parallel. Neither has been chosen, and the economics have not been compared. Each challenge below can be answered on its own; a proposal does not have to solve the whole cell.
Challenges
11 open. Each one can be answered on its own — a proposal does not have to solve the whole cell.
- 1
Getting into the sample vessel
Lid off and back on (what the lab's step diagram shows), a septum lid, or a pierceable membrane? The viscosity of the sample decides what is even mechanically possible, and the choice sets the sterility story for everything downstream.
Interfaces
Receives the sample's boiled and unboiled portions from the dissolution step; hands a dosed volume to the filtration unit.
- 2
Disinfection between samples
The lab has narrowed it down: UV alone leaves shadowed surfaces untreated, hydrogen peroxide is expensive, and flammable sprays need extraction so no explosive atmosphere forms. Still open: hot air (what the TU Dresden demonstrator uses), steam sterilising in place at 121 °C under overpressure, ozone broken down afterwards over activated carbon, infrared or ultrasound. Which fits the cycle time, and does a manual ethanol wipe before a run remain necessary?
Interfaces
Owns every surface shared between samples — grippers, frits, funnels, filtration units — and must hand the next sample a clean path.
- 3
Frit handling
One frit carries all seven membranes of a sample. After the sample: swap it for a sterile one, disinfect it in place, or move the whole rack out to be sterilised while a second rack takes over? Could magnetic frits make handling easier, and can frits be cleaned automatically? Which option is cheaper to build and to run has not been established.
Interfaces
Presents a sterile frit to the filtration unit at the start of each sample and takes the used one away for cleaning or sterilising.
- 4
Transferring the membrane to the plate
Lift a wet membrane off the frit and lay it flat onto agar without folding, tearing or trapping air — then close the plate and keep its identity attached. Seven times per sample, at about 40 seconds per filtration on average.
Interfaces
Takes a wet membrane from the filtration unit; hands a closed, identified plate to the stacking step.
- 5
Presenting the consumables
Funnels, plates, membranes (possibly of two types), tips and frits all have to arrive sterile and on demand: about 700 membranes, plates and tips a day. Carousel, stack magazines, or flexible stacking systems — and how much buffer is needed to reach the walk-away time.
Interfaces
Accepts a manual sterile restock before a run; presents each consumable to the robot on demand and reports when one runs out.
- 6
Aborting a filtration that stalls
Rare with white sugar, but it needs a concept. After 10 minutes: alarm, then either (A) break the filter and let the vacuum pull the liquid through, or (B) pipette the remaining liquid off, automatically or by hand. Either way the filter is discarded, the frit is cleaned, a sterile frit goes in, and the run continues.
Interfaces
Takes over when the pressure or liquid sensor reports no progress; returns the station to a state the next sample can use.
- 7
Handover to incubation
Finished plates are stacked by incubation condition (48 h at 55 °C, 48 h at 30 °C, 72 h at 30 °C, 48 h at 36 °C) and handed to the incubators, automatically or by hand, with the sample-to-plate mapping recorded.
Interfaces
Receives closed plates carrying their sample identity; hands trays to incubators running at 30 °C, 36 °C and 55 °C.
- 8
Dosing 3 to 40 ml without under-dosing
Sample volumes range from 3 to 40 ml and must land within ±10 %; under-dosing is worse than over-dosing. Standard robotic tips stop well below 40 ml, so it is open whether larger tips exist or whether a different dosing method is needed. A fresh sterile tip is used for each dose.
Interfaces
Draws from the opened sample vessel; doses into a funnel that already holds 100 ml of sterile water.
- 9
Choosing the robot architecture
An arm with a tool changer (gripper, suction, pipette, membrane handling) or a gantry with fixed tools and no tool changes? The arm is more flexible; the gantry makes parallel work easier. Neither has been costed.
Interfaces
Has to reach the consumable stores, the sample hand-over, every filtration position, the plate stacks and the waste.
- 10
How many filtrations run in parallel
A sample needs 7 filtrations and the budget is about 40 seconds per filtration, so several must run at once. The seven of one sample share a frit and run in order, so parallel positions serve different samples. Strips of 2, 5 or 6 positions? And does each vacuum position need its own valve and sensor so positions can start and finish independently?
Interfaces
Shares the vacuum source and waste line across positions; each position reports its own end-of-filtration signal.
- 11
Sterile hand-over from upstream
Dissolved samples, and possibly freshly poured plates, arrive from upstream steps that may be automated later. How do they enter the enclosure without breaking its sterility?
Interfaces
Upstream: sample dissolution, media preparation, plate pouring. Inside: the sample-access and plate-presentation steps.
What the figures imply
- 700 of everything. 700 filtrations, membranes, agar plates and pipette tips a day, plus the negative controls.
- About 150 litres a day through the system. 140 litres of sterile process water (200 ml per filtration) and about 11 litres of sample, all of it into liquid waste.
- About 106 ml of each sample is used: 30 ml of the boiled portion and 76 ml of the unboiled one, before any dead volume.
- Several positions must run at once. At about 40 seconds per filtration, a filtration that occupies a position for 2 minutes needs 3 positions running side by side; 4 minutes needs 6. The seven filtrations of a sample share one frit and run in order, so the parallel positions serve different samples.
- About 25 samples on board. 120 minutes of walk-away at 4.8 minutes per sample is 25 samples: 175 membranes, plates and tips, 25 frits, and 25 to 175 funnels depending on whether a funnel may be reused within a sample.
What a useful proposal covers
A concept does not have to answer everything at once, but it should say how it handles:
- the station layout and robot variant, with a cycle-time estimate against about 40 seconds per filtration and 700 a day
- the order rule: seven membranes per sample on one frit, boiled first
- moving the membrane from frit to agar without folding, tearing or trapping air
- disinfection between samples: which method, how long it takes, and how it would be validated
- how many consumables it holds, and how that reaches 120 minutes of walk-away
- dosing 3 to 40 ml within ±10 % without under-dosing
- stalled-filtration recovery and the traceability record
- single-use versus reusable parts, with running costs
More media
Figure 9
Submit a solution
Have an idea how to solve this, or a part of it? Send it to Birdwave: a comment, and a document if you have one — a concept, a layout, a datasheet. Only Birdwave sees what you send: not the company, not other members.
Sign in to submit a solution.
Sign inQuestions & comments
Sign in to ask a question about this use case, or to read and join the discussion.
Sign in





